3)Warm Water thermal effect
Low High
Adjustment
Shaft
[5]Operation 6.
If warm air is not needed in the summer or at any other time, disconnect the
heater inside the tube to turn it off.
>>If the heater gets too hot, the heater and the tube could be damaged, and the
working life of both will be shortened.
If the three-way syringe tube is equipped with a heater, the air it
emits can be warmed.
Turn off the main switch, and then follow the procedure below. (Same
for both the dentist’s and the assistant’s syringe.)
a)Plug the tube connector into the chair connector(at the side of where
the main tube is connected to the chair).
b)Turn on the main switch to let the heater inside the tube work.
Let the warm water spout and then check its temperature.
<Temperature Adjustment>
Use a miniature screwdriver to turn the temperature adjustment shaft
(see diagram to the right).
The dentist’s and assistant’s syringes are adjusted separately.
How warm the air gets depends on the room
temperature and how frequently it is used.
Make sure the heater is properly adjusted and
does not overheat.
Temperature adjustment for dentist’s side three-way syringe:
Turn the adjustment shaft on the bottom of dentist’s table.
Temperature adjustment for assistant’s side threeway syringe:
Turn the adjustment shaft in the basin unit.
Chair Connector
Tube Connector
Adjustment Shaft
*:option
CHECK
CAUTION