Chapter Five: Diagnostic Interface Setup, Configuration,
and Operation
Monitor Mode
49
To run this test select, select the “Temperature:” checkbox, and then press the “Run the
Test” button. When the test finishes, the test result “Pass” or “Fail” will be displayed next
to the checkbox. If the test passes, then the temperature circuit is good. If it fails, then there
may be a problem.
Figure 17: IE1000A Embedded Interface (Monitor Mode / Diagnostics Page)
Monitor Mode / Configuration Page
This Monitor Mode / Configuration Page displays the TCP/IP settings and the current firmware version for
the MFC. The Configuration Page allows the user to enter the “Setup Mode” of the device when the proper
access password is supplied. The factory-shipped password is “config” (without the ““). The figure below
shows a screen capture of the Configuration Page in “Monitor Mode.”