Step 2: Configure TCP/IP Settings for Communication to
IE1000A
Chapter Five: Diagnostic Interface Setup, Configuration, and
Operation
44
Figure 13: Windows 7 LAN Properties Dialog Box
5.
Left click to highlight <Internet Protocol Version 4 (TCP/IPv4)>. Select <Properties>
6.
A new pop up will appear related to “Internet Protocol Version 4 (TCP/IPv4) Properties” which allows
you to change the default IP address of the communication port:
Figure 14: Windows 7 IP Properties Dialog Box
7.
Select “Use the following IP Address”
8.
Enter a unique IP address.
Note
The IE1000A High Flow Rate Mass Flow Controller’s IP address format is 192.168.2.155. Your
host computer’s IP address must also be of the format 192.168.2.XXX, where XXX is some
number between 0 and 255. The host computer IP address must be unique. A recommended IP
address for your Host computer is 192.168.2.2 with a subnet mask of 255.255.255.0.