Device Installation
PR 5500 Instrument Manual
EN-46 Sartorius
4.4.1.2 Connecting to a PLC or RS-485/RS-232 Converter
Point-to-point connection for the ModBus protocol (2-wire)
Example
PR 5500/04
RS-485-A
PR 5500
A-TxB
RxD/TxD-P
RxD/TxD-N
A-TxA
A-RxB
A-RxA
GNDI
+ 24 V
GND
DC
DC
1
2
3
4
5
6
7
8
↓
ON
A
{
S4
120
S1
S3
120
S2
1K6
1K6
A
PLC
ModBus-RTU (master)
TxB/RxB
TxA/RxA
GND
Switch settings
ON:
S2, S3, S4
OFF:
S1
Configuration
[System setup] – [Connected devices] – [ModBus-RTU master] – [Interface] – [Option-1 RS-485-A]