Commissioning
PR 5500 Instrument Manual
EN-212 Sartorius
5.14.10 Corner Correction
5.14.10.1 Checking Corner Load (Dead Load)
After assignment and calibration, the load cell positions have been defined clearly.
5.14.10.2 Mechanical Corner Correction
A mechanical corner correction must carry out if the load on the load cells is not evenly distributed, e.g when
the platform wobbles. Now the individual dead load on the load cells can be corrected using shims. If two
coupled platforms are connected, corner load checking or installation of shims for the platforms must be
performed independently.
Access the Pendeo setup menu as follows: [System setup]-[Weighing point]-[Weighing point A]-[Calib].
t
t
Select and confirm the menu item ‘View and
assign load cells’.
t
y
The load cells are displayed with their item
number, serial number, and load.
t
t
In this example, the load on a load cell doesn’t
need to be increased by a shim.
t
t
A fine correction can be conducted with a
software corner correction, see Chapter 5.14.10.3.
t
t
Press the ESC/EXIT key to return to the previous
window.