PR 5500 Instrument Manual
Commissioning
Sartorius EN-217
5.14.12 Subsequent Dead Load Correction
If the platform weight changes by an amount that is higher than the zero-setting range, e.g., due to dead load
reduction, dead load increase, or mechanical changes, the functions for automatic zero tracking and manual
zero setting no longer work.
If the entire zero-setting range is already utilized, you can still correct the dead load without affecting other
calibration data/parameters.
Access the menu as follows:
[System setup]-[Weighing point]-[Weighing point A]-[Calib]-[Calibrate the scale]-[Modify].
The scale must be empty (deck installed on a vehicle
scale, no load).
t
t
Select ‘Dead load’.
t
t
Confirm the dead load (
s
corner load).
t
t
Check the corner load, see Chapter 5.14.10.1.
t
t
Carry out a corner correction if necessary, see
Chapter 5.14.10.
t
t
Press the [Save] soft key to save the current dead
load.