Commissioning
PR 5500 Instrument Manual
EN-280 Sartorius
t
t
Confirm the line.
t
y
An information windows appears.
[Zero correction]
The zero point correction in use is displayed.
[Communication error count]
The communication errors (time frames exceeded)
for the load cells are counted here in ascending
order and displayed.
[WZ 1…n]
Bar graph display
A
B
C
The bar graph shows three areas:
A = Dead load (can be changed by calibration)
B = Nominal load E
max
(max. capacity of load cell) including dead load (load cell, cannot be changed)
C = Max. usable load including dead load (load cell, cannot be changed)
The colors have the following meanings:
-
Red: Weight value is above maximum capacity (overload) or below -¼ d.
-
Green: Weight value is within tolerances.
-
Orange: Weight value is above nominal load E
max
(max. capacity of load cell).