PR 5500 Instrument Manual
Commissioning
Sartorius EN-177
[Standstill time]
The parameters [Standstill time] and [Standstill range]
can be used to define the stability of the scale (stable
balance).
The [Standstill time] parameter is entered in seconds.
The permitted range is 0.00…2 s.
The time can be entered from 0.00 to 2.00 seconds,
but makes up at least one measurement time.
[Standstill range]
As long as the weight fluctuations remain within this range, the device is determined to be stable.
The [Standstill range] parameter is entered in “d.” The permitted range is 0.01…10.00 d.
[Tare timeout]
Timeout for a tare/zeroset command that cannot be executed (e.g., due to mechanical instability of the scale,
incorrect filter setting, resolution too high, standstill condition too strict).
This parameter is given in seconds. The permitted range is 0.0…<2.5>…25 s.
At 0.0 s taring is only carried out when the scale is already stable.
[Zeroset range]
Determine a ±range around the zero point determined by the dead load during calibration; within this range
-
the displayed gross weight can be set to zero by pressing the zero-setting key (or by a corresponding
external command), and
-
automatic zero tracking is active.
Setting range: 0.00…10000.00 d
[Zerotrack]
The zero display is automatically maintained within set limits.
Possible values: no/yes
When “No” is selected, the next three parameters are not shown.
When “Yes” is selected values for the next three parameters must be entered.
[Zerotrack range]
Range within which the automatic zerotrack is balanced.
Setting range: 0.25…10000.00 d