Commissioning
PR 5500 Instrument Manual
EN-116 Sartorius
t
y
The icons for the devices are displayed.
t
t
Click on the relevant icon using the right mouse
button and select the ‘Properties’ menu item.
t
t
Read the IP address.
UPnP View with Microsoft Windows 7
If the network was
not
configured as a ‘public network’, the device icons are shown automatically under
‘Network’.
5.6.2 Activating the ‘DHCP’ Network
Requirement
‘Use DHCP’ has been selected under [System setup]-[Network parameters].
This ensures that a valid address for device network identification can be assigned to the device;
see Chapter 5.19.4.
Note
The last three bytes of the MAC ID are displayed in default host name. A label with the
complete MAC ID is located on the outside of the device.
A device with DHCP ‘on’ (default factory setting) and connected to an IT network (company
network) with a DHCP server does not require any further action, but there will be a
delay of 2–3 minutes. A network connection is then established automatically (device <->
workstation/PC).
Notebooks/PCs connected temporarily must have the following network adaptor properties (DHCP/DNS
automatic):
For setup, the first network contact is possible only by determining the IP address/subnet mask automatically
under ‘Obtain an IP address automatically’ (default factory setting) or ‘Use the following IP address’.