Commissioning
PR 5500 Instrument Manual
EN-154 Sartorius
5.11 Calibrating the Internal Weighing Point
5.11.1 General Information
With applications for use in legal metrology, the legal requirements and the conditions given on the test/
approval certificate must be taken into account when selecting the settings.
There are two ways to save the calibration data:
-
Via a CAL switch
-
Via software
During calibration, the instrument must be set to gross weight display (reset tare, if necessary).
5.11.1.1 CAL Switch
The calibration data can be protected with CAL switch A and the weighing electronics with CAL switch 1 (see
Chapter 5.1.5.1).
5.11.1.2 Software
A unique check number is created every time a calibration or changed parameters are saved. This can be viewed
under [System information]-[Show calibration check numbers] (see Chapter 5.20.7) and noted.