PR 5500 Instrument Manual
Device Installation
Sartorius EN-103
Link/Activity
LED status
Description
Comments
Off
Module has no connection
-
No connection
-
No communication
Constant green
Module has a connection
-
There is an Ethernet connection
-
No communication
Flickering green
Activity
-
There is an Ethernet connection
-
Communication is available
4.4.11.3 Fieldbus parameters
Slave/Master device names
Caution!
A unique device name must be assigned from the master. This name is given highest
priority when establishing the communication.
When replacing devices/servicing, please note:
Along with the IP address, the device name must correspond to the one of the
replacement device. Explicit assignment out of the master is required.
Example
Slave
Slave
Master
PLC
PR5500-Scale1
PR5500-Scale2
PR5500-Scale2
PR5500-Scale1
Device 1
Device 2