PR 5500 Instrument Manual
Extended Functions
Sartorius EN-305
t
y
The window reappears, now displaying the
sequence number for the selected data record (in
this case, 10) under [First sequence number].
t
t
Repeat the process to choose the last sequence
number.
t
t
Select the output parameters.
[Compression]
Selection:
<no compression>,
Best speed (reduced size by a factor of ~10,
requires ~10 % longer),
Best compression (reduced size by a factor of ~20,
requires ~70 % longer)
[Destination]
Selection:
USB stick,
network share connections (configured under
[System setup] - [Network share connections])
t
t
Press the [Export] soft key to start the export.
t
y
A prompt window appears.
t
t
Press the [Continue] soft key.
Note
If no USB stick is plugged in, a warning appears.
t
t
Plug in a USB stick and wait until the symbol appears in the info line.
t
t
Press the [Retry] soft key.
t
y
The XML file is saved to the USB stick.
t
y
As soon as the process is complete, an info
window appears.
A progress window appears during the save
process.
t
t
Press the [OK] soft key to return to the [Alibi
memory maintenance] menu.