PR 5500 Instrument Manual
Device Installation
Sartorius EN-99
4.4.10.2 LEDs
1
5
ERR
RUN
CC-Link
Pos.
Description
RUN
Process LED
ERR
Error LED
Operation (1)
LED status
Description
Comments
Off
Module is offline
-
No network connection
-
No power
Constant green
Module is online
-
There is a network connection
-
Normal operation
Constant red
Exception error
-
Module is in the EXCEPTION
status
-
The exception error monitoring
system closes all open
connections to the module.
Error (2)
LED status
Description
Comments
Off
No error
No power
Constant red
Exception error
-
Module is in the EXCEPTION
status
-
The exception error monitoring
system closes all open
connections to the module.
Flickering red
CRC error
Flashing 1 Hz red
-
Address error
-
Baud rate error
After any changes:
-
No allowed addresses
-
No allowed baud rates