PR 5500 Instrument Manual
Commissioning
Sartorius EN-189
5.12.5 Setting the xBPI Dead Load
Note
Both terms ‘dead load’ and ‘preload’ are used by Sartorius.
t
t
Press the [Calib] soft key to calibrate the
xBPI platform.
t
y
The PR 5500 is reading the parameters from the
xBPI scale.
t
y
Ticks indicate the progress.
t
y
An error message appears if communication with
the xBPI scale is not possible!
Setting the Dead Load
t
t
Remove the weight from the scale.
t
t
Select and confirm [Set] using the cursor.
t
y
After the command is sent, the gross weight
display shows "0".