PR 5500 Instrument Manual
Commissioning
Sartorius EN-225
t
t
Select the line ‚Number of vessel feet‘, change if
necessary and confirm.
t
t
Select the line ‚Local gravity‘, change if necessary
and confirm.
t
t
Enter the maximum capacity with decimal places
(in this example: 20.000 t).
t
t
Double-click on the toggle key opens a small
selection window.
t
t
Select the unit with cursor keys
/
and confirm
with OK key.
t
t
Confirm the entries.
t
y
The confirmation is displayed with ‘setting Max...’.
The weight unit is taken from [Max]. The number of
digits behind the decimal point is also automatically
determined when [Max] is entered.
t
t
Select the scale interval (1 d).
t
y
The scale interval (d) is then calculated, based on
the Max weight value.
t
t
Confirm the entries.
t
y
The confirmation is displayed with ‘setting scale
interval...’.
Содержание Sartorius Intec Maxxis 4 PR 5500
Страница 406: ......