Specifications
PR 5500 Instrument Manual
EN-384 Sartorius
12.3.5 Power Connection 24 V DC
Supply voltage
24 V DC
±10%
Max. power consumption
1 W
Primary fuse
1x 2 AT, 250 V, 5 x 20 mm
e.g. made by Schurter: SPT5x20, order no.: 0001.2507
12.4 Effect of Ambient Conditions
12.4.1 Ambient Conditions
Temperature range
Ambient temperature for operation
-10 to +50°C
Ambient temperature for use in legal metrology
-10 to +40°C
Power-on temperature
> 0°C
Storage/transport
-20 to +70°C
Moisture
< 95%, no condensation
(in accordance with IEC 68-2)
Protection class
IP65 front, IP20 rear
Elevation
< 2000 m
12.4.2 Electromagnetic Compatibility (EMC)
All data in compliance with NAMUR NE 21, EN 45501, and EN 61326.
Housing
Radio frequency electromagnetic fields
(80-3000 MHz)
EN 61000-4-3
10 V/m
Electrostatic discharge (ESD)
EN 61000-4-2
6/8 kV
Signal und
Fast transients (bursts)
EN 61000-4-4
1 kV
control lines
Peak voltages (surges) 1.2/50 µs
EN 61000-4-5
1/2 kV
Conducted disturbances by radio frequency
(0.15-80 MHz)
EN 61000-4-6
10 V
Mains inputs
Fast transients (bursts)
EN 61000-4-4
2 kV
Peak voltages (surges) 1.2/50 µs
EN 61000-4-5
1/2 kV
Conducted disturbances by radio frequency
(0.15-80 MHz)
EN 61000-4-6
10 V
Voltage dips
EN 61000-4-11
0/40/70%
20/200/500 ms
Mains failure link
EN 61000-4-11
20 ms