PR 5500 Instrument Manual
Commissioning
Sartorius EN-257
[Screensaver]
Once the screensaver is switched on, operation is only possible once a password has been entered or a key
has been pressed.
Selection:
[no screensaver], after 1 minute, after 5 minutes, after 10 minutes, after 30 minutes
[Keyclick sound]
Duration of acoustic signal when keys are pressed
Selection:
no key click, <short key click>, medium key click, long key click
[Coldstart with STOP+EXIT]
A cold start leads to a loss of all data on the database.
!
Caution
Selection:
<disabled>, immediately, 3 s
[Software download]
Load and start a new application on the device via the network.
Possible production downtime!
This parameter must always be set to “locked” for production systems.
!
Caution
Selection:
<disabled>, enabled
Software can only be downloaded in running operation when “enabled” has been selected.
[Label/Language download]
Load labels and language files via network temporarily into the device.
Note
After a cold start the files will be lost.
Loading the language files permanently into the device is possible via the update
software “FlashIt!” from version 02.73.11.
Selection:
<disabled>, enabled
[Keys require logged in user]
If the checkbox
is ticked, the keys with weighing point functions (e.g. Set zero; F1 and F2, if
configurated for weighing point functions) are locked for users who are not logged in.