PR 5500 Instrument Manual
Commissioning
Sartorius EN-151
5.10.13 System setup: Weighing point: Weighing point A: PR-Net Weighing Point
System setup
— Weighing point
— Weighing point A
PR-Net weighing point
— Parameters
Parameter settings
— Settings locked
When selected settings are displayed only. No changes possible (as when
CAL switch is closed).
— W&M
<none>, [OIML], NSC, NTEP
— Serial number
<0>…999999999
— Network address
Numeric input:
IP address (nnn.nnn.nnn.nnn)
or
Alphanumeric input:
Device name (host name)
— Default
Settings are reset to factory settings.
— Save
The settings are saved.
5.10.14 System setup: Weighing point: Weighing point A: Mettler-Scale
System setup
— Weighing point
— Weighing point A
Mettler-Scale
— Param
Parameter settings
— Settings locked
When selected settings are displayed only. No changes possible (as when
CAL switch is closed).
— W&M
<none>, [OIML], NSC, NTEP
— Max (Maximum capacity)
0.000001 mg…<1000 kg>…9999999 t
— Scale interval
<1>, 2, 5, 10, 20, 50
— Weight per
0.000001 mg…<1 kg>…9999999 t
— Minimum weight
0 d…9999999 d
— Update time
0.1 s…<0.3 s>…2.0 s
— Default
Settings are reset to factory settings.
— Save
The settings are saved.