Commissioning
PR 5500 Instrument Manual
EN-148 Sartorius
5.10.12 System setup: Weighing point: Weighing point A: Pendeo Process
System setup
— Weighing point
— Weighing point A
Pendeo Process
— Parameters
Parameter settings
— Settings locked
When selected settings are displayed only. No changes possible (as when
CAL switch is closed).
— W&M
<none>, [OIML] (not possible when range mode "multi-interval" is
selected or Max has been set with more than three decimal places), NSC,
NTEP
— Ambient conditions
very stable condition, Stable condition, Unstable condition, Very
unstable condition
— Detect a defect load cell
Select
if max. simulation of a faulty load cell should be automatic.
— Unbalanced check
deviation
The plausibility check is activated when the average deviation is >0%.
Setting range: 0–100%
— Standstill time
0.00 s…<0.50 s>…2.00 s (The range depends on the measurement time.)
— Standstill range
0.00 d…1.00 d…10.00 d (The range depends on the measurement time.)
— Tare timeout
0.0 s…<2.5 s>…25.0 s; timeout when there is no standstill
— Zeroset range
0.00 d…<50.00 d>…10000.00 d [<4% and <20% of Max],
timeout duration when there is no standstill
— Zerotrack
<yes>, no; when "yes" is selected values for the next three parameters
must be entered.
— Zerotrack range
0.01 d…0.25 d…10000.00 d (only if "Zerotrack" is set to "yes")
— Zerotrack step
0.01 d…0.25 d…10.00 d (only if "Zerotrack" is set to "yes")
— Zerotrack time
0.1 s…1.0 s…25.0 s [<0,5 e/s] (only if "Zerotrack" is set to "yes")
— Overload
0 d…<9 d>…9999999 d [
<
9 e]
— Minimum weight
0 d…<20 d>…9999999 d [20 e for class III]
— Range mode
<single range>, multiple range, multi-interval
See also Chapters 5.11.16.2 and 5.11.16.3.
— Range limit 1
0…max. weight, transition from small to medium scale interval; only
when multiple range or multi-interval has been selected.
— Range limit 2
0…max. weight, transition from medium to large scale interval; only
when multiple range or multi-interval has been selected.
— Default
Settings are reset to factory settings.
If “Settings locked” is selected, the [Default] soft key is not active.
— Save
The settings are saved.