3-10
WARNING
Do not hot plug T1 dock data cable. Hot plug may result in unknown problems.
Make sure that the T1 dock data cable is disconnected from the host monitor when T1 docking station is
not in use with the host monitor.
3.11.7
Demo Mode
In Demo mode, the monitor can demonstrate its major functions when patient or patient simulator is not connected.
The Demo mode is password protected.
To enter the Demo mode,
1. Select
[
Main Menu
]
→
[
Maintenance >>
].
2. Select
[
Demo >>
]. Enter the required password and then select [
Ok
].
To exit the Demo mode, select [
Main Menu
]
→
[
Maintenance >>
]
→
[
Exit Demo
]
→
[
Ok
].
WARNING
The Demo mode is for demonstration purpose only. To avoid that the simulated data are mistaken for the
monitored patient’s data, you must not change into Demo mode during monitoring. Otherwise, improper
patient monitoring and delayed treatment could result.
3.11.8
Standby Mode
In standby mode, you can temperately stops patient monitoring without turning off the monitor. To enter the standby
mode, select [
Main Menu
]
→
[
Standby
].
Содержание BeneView T1
Страница 1: ...BeneView T1 Patient Monitor Operator s Manual...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES...
Страница 20: ...12 FOR YOUR NOTES...
Страница 38: ...2 14 FOR YOUR NOTES...
Страница 56: ...4 8 FOR YOUR NOTES...
Страница 64: ...5 8 FOR YOUR NOTES...
Страница 116: ...8 30 FOR YOUR NOTES...
Страница 130: ...11 8 FOR YOUR NOTES...
Страница 138: ...12 8 FOR YOUR NOTES...
Страница 140: ...13 2 FOR YOUR NOTES...
Страница 174: ...16 12 FOR YOUR NOTES...
Страница 190: ...18 10 FOR YOUR NOTES...
Страница 206: ...21 6 FOR YOUR NOTES...
Страница 246: ...A 20 FOR YOUR NOTES...
Страница 274: ...D 8 FOR YOUR NOTES...
Страница 285: ...G D Declarat tion of C Conform G 1 mity...
Страница 286: ...P N 046 005297 00 15 0...