23-4
23.5
NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required at least once a year or when you
doubt the measured NIBP. If the test failed, corresponding prompt messages will be given. If no message is displayed, it
means no leakage is detected.
Tools required:
An adult cuff
An air tubing
A correct sized cylinder
Follow this procedure to perform the leakage test:
1.
Set the patient category to [
Adu
].
2.
Connect the cuff to the NIBP connector on the monitor.
3.
Wrap the cuff around the cylinder as shown below.
4. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
User
Maintenance >>
]. Enter the required password and then select
[
OK
].
5. Select
[
Module Maintenance >>
]
→
[
NIBP Leakage Test
]. The NIBP display shows [
Leakage Testing…
].
After about 20 seconds, the monitor will automatically deflate. This means the test is completed. If the message [
NIBP
Pneumatic Leak
] is displayed, it indicates that the NIBP airway may have leakages. Check the tubing and connections
for leakages. If you ensure that the tubing and connections are all correct, perform a leakage test again.
If the problem persists, contact your service personnel.
Monitor
Connector for
NIBP cuff
Air tubing
Cylinder
Cuff
Содержание BeneView T1
Страница 1: ...BeneView T1 Patient Monitor Operator s Manual...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES...
Страница 20: ...12 FOR YOUR NOTES...
Страница 38: ...2 14 FOR YOUR NOTES...
Страница 56: ...4 8 FOR YOUR NOTES...
Страница 64: ...5 8 FOR YOUR NOTES...
Страница 116: ...8 30 FOR YOUR NOTES...
Страница 130: ...11 8 FOR YOUR NOTES...
Страница 138: ...12 8 FOR YOUR NOTES...
Страница 140: ...13 2 FOR YOUR NOTES...
Страница 174: ...16 12 FOR YOUR NOTES...
Страница 190: ...18 10 FOR YOUR NOTES...
Страница 206: ...21 6 FOR YOUR NOTES...
Страница 246: ...A 20 FOR YOUR NOTES...
Страница 274: ...D 8 FOR YOUR NOTES...
Страница 285: ...G D Declarat tion of C Conform G 1 mity...
Страница 286: ...P N 046 005297 00 15 0...