E-4
E.7
Mains on Applied Part Leakage
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage, through a limiting resistance,
to selected applied part terminals. Current measurements are then taken between the selected applied part and earth.
Measurements are taken with the test voltage (110% of mains) to applied parts in the normal and reverse polarity
conditions
The following outlet conditions apply when performing the Mains on Applied Part test.
Normal Polarity;
Reversed Polarity
LIMITS
For CF applied parts: 50 μA
For BF applied parts: 5000 μA
E.8
Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the remaining Applied Part
connector s. All measurements may have a true RMS only response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
normal polarity( Normal Condition);
reverse polarity( Normal Condition),
normal polarity with open neutral(Single Fault Condition);
reverse polarity with open neutral(Single Fault Condition).
normal polarity with open earth(Single Fault Condition);
reverse polarity with open earth(Single Fault Condition).
LIMITS
For CF applied parts,
10μA in Normal Condition
50μA in Single Fault Condition
For BF applied parts,
100μA in Normal Condition
500μA in Single Fault Condition
NOTE
Make sure the safety analyzer is authorized comply with requirement of IEC60601-1.
Follow the instructions of the analyzer manufacturer.
Содержание BeneView T1
Страница 1: ...BeneView T1 Patient Monitor Operator s Manual...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES...
Страница 20: ...12 FOR YOUR NOTES...
Страница 38: ...2 14 FOR YOUR NOTES...
Страница 56: ...4 8 FOR YOUR NOTES...
Страница 64: ...5 8 FOR YOUR NOTES...
Страница 116: ...8 30 FOR YOUR NOTES...
Страница 130: ...11 8 FOR YOUR NOTES...
Страница 138: ...12 8 FOR YOUR NOTES...
Страница 140: ...13 2 FOR YOUR NOTES...
Страница 174: ...16 12 FOR YOUR NOTES...
Страница 190: ...18 10 FOR YOUR NOTES...
Страница 206: ...21 6 FOR YOUR NOTES...
Страница 246: ...A 20 FOR YOUR NOTES...
Страница 274: ...D 8 FOR YOUR NOTES...
Страница 285: ...G D Declarat tion of C Conform G 1 mity...
Страница 286: ...P N 046 005297 00 15 0...