C-14
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
High Alarm Interval (s)
No
Yes
10
Med Alarm Interval (s)
No
Yes
20
Low Alarm Interval (s)
No
Yes
20
Alarm Light on Alarm Reset
No
Yes
On
Reset Other Bed's Alarms
No
Yes
Off
Alarm Reset By Other Bed
No
Yes
On
Minimum Alarm Volume
Yes
Yes
General: 2
OR: 1
ICU, NICU, CCU: 2
Alarm Sound
No
Yes
ISO
Reminder Tone
No
Yes
On
Reminder Interval
No
Yes
3 min
ECGLeadOff Lev.
No
Yes
Low
SpO
2
SensorOff Lev.
No
Yes
Low
IBPSensorOff Lev.
No
Yes
Med
Lethal Arrh. OFF
No
Yes
Disable
Extended Arrh.
No
Yes
Disable
Alarm Delay
No
Yes
6 s
ST Alarm Delay
No
Yes
30 s
Other Bed Disconnection Alm
No
Yes
On
Wave Line
No
Yes
Mediate
Outdoor Mode
No
Yes
Manual
ECG Standard
No
Yes
AHA
Notch Freq.
No
Yes
50 Hz
Data Transfer Method
No
Yes
Off
Transferred Data Length
No
Yes
4 h
Para Switch Authority
No
Yes
Unprotected
Parameter Switch
Yes
Yes
When [
Para Switch Authority
] is set
to [
Protected
]: Unselected
When [
Para Switch Authority
] is set
to [
Unprotected
]: Selected
SpO
2
Tone
No
Yes
Mode 1
Содержание BeneView T1
Страница 1: ...BeneView T1 Patient Monitor Operator s Manual...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES...
Страница 20: ...12 FOR YOUR NOTES...
Страница 38: ...2 14 FOR YOUR NOTES...
Страница 56: ...4 8 FOR YOUR NOTES...
Страница 64: ...5 8 FOR YOUR NOTES...
Страница 116: ...8 30 FOR YOUR NOTES...
Страница 130: ...11 8 FOR YOUR NOTES...
Страница 138: ...12 8 FOR YOUR NOTES...
Страница 140: ...13 2 FOR YOUR NOTES...
Страница 174: ...16 12 FOR YOUR NOTES...
Страница 190: ...18 10 FOR YOUR NOTES...
Страница 206: ...21 6 FOR YOUR NOTES...
Страница 246: ...A 20 FOR YOUR NOTES...
Страница 274: ...D 8 FOR YOUR NOTES...
Страница 285: ...G D Declarat tion of C Conform G 1 mity...
Страница 286: ...P N 046 005297 00 15 0...