A-15
A.8.7
IBP
Standards
Meet standard of IEC60601-2-34.
Technique Direct
invasive
measurement
IBP
Measurement range
-50 to 360 mmHg
Resolution 1
mmHg
Accuracy
±2% or ±1 mmHg, whichever is greater (without sensor)
Refreshing rate
1 s
Pressure transducer
Excitement voltage
5 VDC, ±2%
Sensitivity 5
μV/V/mmHg
Zero adjustment range
±
200 mmHg
Impedance range
300 to 3000Ω
Alarm limit
Range (mmHg)
Step (mmHg)
Sys High
(low limit + 2) to 360
1
Mean High
Dia High
Sys Low
-50 to (high limit – 2)
Mean Low
Dia Low