7-14
You can separately select [
Latching Visual Signal
].
Selecting [
Latching Audible Signal
] simultaneously latches the visual signal.
Selecting alarms of lower priority simultaneously latches the alarms of higher priority.
7.10
Using Care Group Alarms (Only Available for the External Display)
7.10.1
Care Group Auto Alarms
If any monitor in the Care group not being viewed by your monitor is alarming, a flashing symbol will appear beside the
QuickKeys area. The alarm symbol is shown as below.
The background colors of the alarm symbols indicate alarm levels, and are the same as those of the corresponding alarm
messages. If multiple alarms are active in the Care Group, the background color is the same as that of the highest level
alarm message. For more information about the alarm message and background color, see
7.3.3Alarm Message
.
When a patient monitor in the Care Group is disconnected, the flashing symbol is shown as below.
The department and bed label of the alarming monitor appear on the symbols. You can enter the view other patient
window by pressing the symbol.
7.10.2
Resetting the Care Group Alarms
You can reset the alarms presented on the viewed bed by pressing the [
Alarm Reset
] from the current monitor’s [
View
Other Patient
] window. To enable this function:
1.
Select [
Main Menu
]
→
[
Maintenance>>
]
→
[
User Maintenance>>
]
→
enter the required password
→
[
Alarm
Setup>>
]
→
[
Other Bed Alarm Setup>>
].
2.
Set [
Reset Other Bed’s Alarms
] to [
On
].
The alarms presented on the current monitor can also be reset from another monitor viewing this monitor. To do so,
proceed as follows:
1.
In the current monitor, select [
Main Menu
]
→
[
Maintenance>>
]
→
[
User Maintenance>>
]
→
enter the required
password
→
[
Alarm Setup>>
]
→
[
Other Bed Alarm Setup>>
].
2.
Set [
Alarm Reset By Other Bed
] to [
On
].
3.
In the other monitor, select the [
Alarm Reset
] button from the [
View Other Patient
] window.
NOTE
Changing of alarm priority may affect the latching status of corresponding alarm. Please determine if you
need to reset the latching status for the specific alarm when you have changed its alarm priority.
When the alarm system is reset, the latched physiological alarms are cleared.
Содержание BeneView T1
Страница 1: ...BeneView T1 Patient Monitor Operator s Manual...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES...
Страница 20: ...12 FOR YOUR NOTES...
Страница 38: ...2 14 FOR YOUR NOTES...
Страница 56: ...4 8 FOR YOUR NOTES...
Страница 64: ...5 8 FOR YOUR NOTES...
Страница 116: ...8 30 FOR YOUR NOTES...
Страница 130: ...11 8 FOR YOUR NOTES...
Страница 138: ...12 8 FOR YOUR NOTES...
Страница 140: ...13 2 FOR YOUR NOTES...
Страница 174: ...16 12 FOR YOUR NOTES...
Страница 190: ...18 10 FOR YOUR NOTES...
Страница 206: ...21 6 FOR YOUR NOTES...
Страница 246: ...A 20 FOR YOUR NOTES...
Страница 274: ...D 8 FOR YOUR NOTES...
Страница 285: ...G D Declarat tion of C Conform G 1 mity...
Страница 286: ...P N 046 005297 00 15 0...