Restoring the Latest Configuration Automatically
During operation, you may make changes to some settings. However, these changes may not be saved as user
configuration. To prevent the changes from losing in case of a sudden power failure, the patient monitor stores the
configuration in real time. The saved configuration is the latest configuration.
The monitor restore the latest configuration if restarts within 60 seconds after the power failure. And it will restore the
default configuration rather than the latest configuration if restarts 120 seconds later after the power failure. The
monitor may load either the latest configuration or the default configuration if restarts from 60-120 seconds after the
power failure.
6.11
Modifying Password
To modify the password for accessing the [
Manage Configuration
] menu,
1. Select
[
Modify Password >>
] in the [
Manage Configuration
] menu.
2.
Input a new password in the popup menu.
3. Select
[
Ok
].
Содержание BeneView T1
Страница 1: ...BeneView T1 Patient Monitor Operator s Manual...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES...
Страница 20: ...12 FOR YOUR NOTES...
Страница 38: ...2 14 FOR YOUR NOTES...
Страница 56: ...4 8 FOR YOUR NOTES...
Страница 64: ...5 8 FOR YOUR NOTES...
Страница 116: ...8 30 FOR YOUR NOTES...
Страница 130: ...11 8 FOR YOUR NOTES...
Страница 138: ...12 8 FOR YOUR NOTES...
Страница 140: ...13 2 FOR YOUR NOTES...
Страница 174: ...16 12 FOR YOUR NOTES...
Страница 190: ...18 10 FOR YOUR NOTES...
Страница 206: ...21 6 FOR YOUR NOTES...
Страница 246: ...A 20 FOR YOUR NOTES...
Страница 274: ...D 8 FOR YOUR NOTES...
Страница 285: ...G D Declarat tion of C Conform G 1 mity...
Страница 286: ...P N 046 005297 00 15 0...