D-5
Measurement
Alarm message
L
I
Cause and solution
NIBP Illegally Reset
L
A
An illegal reset occurred during NIBP measurement.
Check if the airway is occluded.
IBP
YY Sensor Off
M*
A
Check the sensor connection and reconnect the sensor.
YY Disconnected
H
C
The liquid way is disconneted from the patient, or the
three-way valve is open to the air. Check the
connection of the liquid way, or check the valve is
open to the patient. If the problem remains, contact
the Customer Services Dept. for help.
YY Sensor Fault
M
C
Replace the sensor.
YY Non-Pulsatile
L
A
The catheter may be occluded. Please flush the
catheter.
YY represents an IBP label.
CCO
Invalid/Faulty PiCCO
catheter
L C
Erroneous or invalid catheter is used. Please use the
proper catheter.
TB Sensor Off
L
A
Check the sensor connections.
PiCCO Comm Abnormal
H
A
Abnormal communication occurred between the
PiCCO module and the system. Remove/connect the
module again or restart the machine. If the problem
remains, contact the Customer Services Dept. for help.
PiCCO Comm Err
H
A
Erroneous communication occurred between the
PiCCO module and the system. Remove/connect the
module again or restart the machine. If the problem
remains, contact the Customer Services Dept. for help.
PiCCO Init Err
H
A
An error occurred to the module during the power-on
self-test. Remove/connect the module again or restart
the machine. If the problem remains, contact the
Customer Services Dept. for help.
Inject Temp. Sensor Err
L
C
An error occurred to the injectate temperature sensor
or the sensor cable. Check/replace the sensor or the
sensor cable.
PiCCO Comm Stop
H
A
Remove/connect the module again or restart the
machine. If the problem remains, contact the Customer
Services Dept. for help.
CO
2
CO
2
Sensor High Temp
L
C
Check, stop using or replace the sensor.
CO
2
Sensor Low Temp
L
C
Check, stop using or replace the sensor.
CO
2
Temp Overrange
L
C
The operating temperature of the CO2 module goes
beyond the specified range. After it restores within the
specified range, the module will restart automatically.
CO
2
Airway High Press.
L
C
An error occurred in the airway pressure. Check the
patient connection and patient circuit, and then restart
the monitor.
CO
2
Airway Low Press.
L
C
CO
2
High Barometric Press.
L
C
Check the CO2 connections, make sure that the
monitor application site meets the requirements, and
check for special sources that affect the ambient
pressure. Restart the monitor.
CO
2
Low Barometric Press.
L
C
Содержание BeneView T1
Страница 1: ...BeneView T1 Patient Monitor Operator s Manual...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES...
Страница 20: ...12 FOR YOUR NOTES...
Страница 38: ...2 14 FOR YOUR NOTES...
Страница 56: ...4 8 FOR YOUR NOTES...
Страница 64: ...5 8 FOR YOUR NOTES...
Страница 116: ...8 30 FOR YOUR NOTES...
Страница 130: ...11 8 FOR YOUR NOTES...
Страница 138: ...12 8 FOR YOUR NOTES...
Страница 140: ...13 2 FOR YOUR NOTES...
Страница 174: ...16 12 FOR YOUR NOTES...
Страница 190: ...18 10 FOR YOUR NOTES...
Страница 206: ...21 6 FOR YOUR NOTES...
Страница 246: ...A 20 FOR YOUR NOTES...
Страница 274: ...D 8 FOR YOUR NOTES...
Страница 285: ...G D Declarat tion of C Conform G 1 mity...
Страница 286: ...P N 046 005297 00 15 0...