2 —25
2 —Monitor Setup
Gas Settings
Figure 2-15: Gas Settings
CO
2
UNIT OF MEASURE
The CO
2
numerical data can be displayed as mm/hg, Torr, kPa, or
volume percent.
AGENT TO MONITOR
The primary agent for monitoring must be correctly entered
depending on the monitoring alarm characteristic desired by the user.
The monitor has two modes of agent gas monitoring. The user may
select a specific halogenated agent to be designated as the primary
agent. The user may otherwise set the monitor to automatically
detect and identify the current primary gas of a mixture.
NOTE: If the user manually selects the primary agent for monitoring it
must be correctly entered.
• Always confirm the primary agent selection before use.
Incorrect primary agent setting may result in erroneous limit
alarms. Alarm characteristics of the monitor are altered when
automatic primary agent detection is activated.
• Never substitute a primary agent setting for a different
halogenated agent, or any agent not listed! The agent detection
is specific to the listed gases only.
• The alarm WRONG AGENT appears when the primary agent
(that is manually selected by the physician) does not match the
primary agent detected. The WRONG AGENT alarm is
deactivated when automatic primary agent detection is
selected.
EXIT
CO2
ON
Unit of measure
mmHg
Color
¤
O2
ON
Color
¤
Agent
Agent to Monitor
Auto
Flow Rate
200 ml/min
Flow Mode
Exhaust
Agent Color
¤
N2O Color
¤
<<< BACK
WARNING
!
!
Содержание Veris 8600
Страница 1: ...Page i VerisTM 8600 Vital Signs Monitor Operation Manual 3010796 Revision 0 Date 10 04 PRELIMINARY...
Страница 10: ...This page intentionally left blank...
Страница 24: ......
Страница 34: ......
Страница 122: ......
Страница 142: ......
Страница 152: ......
Страница 164: ......
Страница 170: ......
Страница 190: ......
Страница 202: ......
Страница 212: ......
Страница 216: ......