If the above Troubleshooting tips don’t resolve the problem, see your
bio-medical personnel or contact MEDRAD’s Customer Service.
Trap error or CO
2
malfunction
• Trap is plugged or leaks
• O
2
Cell is loose
Replace trap
Tighten O
2
Cell
O
2
not responding
(CO
2
functioning ok)
• O
2
Cell depleted
Replace O
2
Cell
No sound from
speaker
• Volume turned down
Set volume higher in Configuration
window.
COM 1 serial interface
doesn’t work
• Set to internal printer
Reset monitor to external printer
Symptom
Problem
Solution