3 —Alarms and Messages
Capnometry (CO
2
)
Alarms and Messages
Priority
Description
LOW EtCO
2
Medium
The end-tidal CO
2
value has dropped below the value
set in the ALARMS menu.
HIGH EtCO
2
Medium
The end-tidal CO
2
value has exceeded the value set
in the ALARMS menu.
LOW INCO
2
Medium
The inspired CO
2
value has dropped below the value
set in the ALARMS menu.
HIGH INCO
2
Medium
The inspired CO
2
value has exceeded the value set
in the ALARMS menu.
LOW RESP
Medium
The respiration value has dropped below the value
set in the ALARMS menu. This is generated from the
capnography module.
HIGH RESP
Medium
The respiration value has exceeded the value set in
the ALARMS menu. This is generated from the
capnography module.
Agent Gas
Alarms and Messages
Priority
Description
AGT:NO BREATH
High
No breath detected. Check patient for apnea. Check
sampling device and adjust placement if necessary.
May indicate a leak in the breathing circuit.
AGT:OCCLUSION
Medium
The sampling line or water trap to the Veris monitor is
completely blocked. The Veris monitor attempts to
clear the block by drawing the occlusion into the
water trap. Replace sampling line as necessary.
CO
2
: BENCH FAIL
Low
A hardware failure has been detected. Contact
MEDRAD Service or your local representative.
CO
2
: NO EXHAUST
Low
Scavenging line on the Veris monitor is blocked or the
scavenging system is defective. Remove blockage or
correct gas scavenging system.
AGT:IR FAIL
Low
A hardware failure has been detected. Contact
MEDRAD Service or your local representative.
AGT:PNEUMATICS
Low
A hardware failure has been detected. Contact
MEDRAD Service or your local representative.
O
2
:SENSOR
Low
The O
2
cell inside the monitor is expended and must
be replaced. Contact MEDRAD Service or your local
representative.
REPLACE TRAP
Medium
The Veris monitor trap needs to be replaced.
AGT: BAD CAL
Low
The Veris monitor was unable to calibrate agent gas
detector. Contact MEDRAD Service or your local
representative.
Содержание Veris 8600
Страница 1: ...Page i VerisTM 8600 Vital Signs Monitor Operation Manual 3010796 Revision 0 Date 10 04 PRELIMINARY...
Страница 10: ...This page intentionally left blank...
Страница 24: ......
Страница 34: ......
Страница 122: ......
Страница 142: ......
Страница 152: ......
Страница 164: ......
Страница 170: ......
Страница 190: ......
Страница 202: ......
Страница 212: ......
Страница 216: ......