MEDRAD
Veris
8600
8 —4
• Do not place the IBP transducer in the bore of the magnet. This
could result in degraded measurement accuracy. Ensure the
transducer is outside of the magnet bore.
• Place only the disposable tubing portion of the IBP kit in the
magnet bore to ensure patient safety and improve measurement
accuracy. Use a disposable set which is a minimum of four feet
(1219 mm) in length.
• Calibrate (zero), the transducer when changing or modifying
transducer tubing.
• When changing or modifying a transducer tubing set, ensure
that the entire fluid path is free of air bubbles. Do not allow air
bubbles to enter the the catheter.