8 —7
8 —IBP
3. Connect the IBP interface cable to the Veris monitor.
4. Connect the appropriate IBP interface cable to the transducer.
5. The monitor displays the message NOT ZEROED in the
informational alarm line. The message appears independently
for each channel.
NOTE: If a waveform slot is set to display IBP, the monitor
shows a waveform immediately. This waveform is representative
of the physiological patient data, but is not yet scaled until the
IBP lines are properly zeroed. No numerical values are
displayed for a channel until it has been zeroed.
6. Select PARAMS and open the first window of parameters.
Figure 8-3: Setting IBP in the PARAMS Window
NOTE: The example window shown contains factory default
values. Reset the parameters as required for actual monitoring
conditions.
7. Verify that the channels are set to the correct IBP sites.
8. Verify that the pressure range is set correctly.
EXIT
HR Source
Smart
Gating
OFF
ECG
ON
Cable
5 lead
Filter
Monitor
Display Range
MEDIUM
Auto lead switching
YES
Color
¤
IBP 1 SITE
ART
RANGE
0 TO 200
ZERO
None
Color ¤
IBP 2 SITE
CVP
RANGE
0 TO 20
ZERO
None
Color ¤
NIBP
ON
GASES
ON
NIBP tone
NONE
Color
¤
ECG/IBP/SPO2 Tone Vol
5
Other Params Menus
No Action
Содержание Veris 8600
Страница 1: ...Page i VerisTM 8600 Vital Signs Monitor Operation Manual 3010796 Revision 0 Date 10 04 PRELIMINARY...
Страница 10: ...This page intentionally left blank...
Страница 24: ......
Страница 34: ......
Страница 122: ......
Страница 142: ......
Страница 152: ......
Страница 164: ......
Страница 170: ......
Страница 190: ......
Страница 202: ......
Страница 212: ......
Страница 216: ......