Procedure for Gas Monitoring
1. Check the Flow Rate settings in the PARAMS softkey window.
2. Slide the WaterChek
™
water trap into the gas receptacle of the
monitor.
3. If anesthetic agents are being used, attach a scavenging line to
the exhaust port at the pedestal of the monitor. A waste gas
recovery system should be used.
4. Attach a sample line to the Luer fixture located on the front of
the WaterChek
™
water trap.
5. Make sure the sampling line is not kinked or pinched.
6. Attach a patient sampling device to the sampling line. Use either
a nasal cannula, mask or ventilation tube adapter.
7. Replace sampling devices, lines, and water traps if they become
blocked or filled.
Occlusions
The monitor displays a visual message AGT: OCCLUSION if the gas
sampling system is blocked. If the sampling line or water trap
becomes partially blocked, the item should be replaced.
Anesthetic Gas
Exhaust Recovery
Always use an exhaust gas recovery system when using anesthetic
gasses. A scavenging kit can be used to connect the monitor to a gas
scavenging system.
1. Slide the tube over the nozzle of the exhaust port located on the
back of the monitor base, near the DC power cord.
2. Connect female Luer-lock end to an exhaust recovery system.
A straight adaptor 19mm O.D./19mm I.D. with a male Luer-lock
entrance is provided with the scavenging kit to provide a connection
to scavenging tubing.
The monitor displays AGT: NO EXHAUST if the scavenging line is
blocked.
Содержание Veris 8600
Страница 1: ...Page i VerisTM 8600 Vital Signs Monitor Operation Manual 3010796 Revision 0 Date 10 04 PRELIMINARY...
Страница 10: ...This page intentionally left blank...
Страница 24: ......
Страница 34: ......
Страница 122: ......
Страница 142: ......
Страница 152: ......
Страница 164: ......
Страница 170: ......
Страница 190: ......
Страница 202: ......
Страница 212: ......
Страница 216: ......