MEDRAD
Veris
8600
H-2
Fuse Specifications
Remote Display Fuses
Number:
2
Type:
T, Slow acting
Amps:
1 amp
Voltage:
250 volts
Main
Monitor
Fuses
NOTE: Fuses in the main monitor are not externally accesible.
Replacement of fuses must be done by a qualified service technician
and the unit must be tested afterwards. Refer to the Service Manual
for fuse replacement. The fuses are found on the Main PCB:
F1:
3A Time Delay, carries 15mA
F2:
4A Time Delay, carries 3A peak
F3:
3A Time Delay, carries 15mA
F4:
1A Time Delay, circuit not used in Veris
F5:
5A Time Delay, carries 4.5A peak
F6:
4A Time Delay, carries 2.5A peak
Power Supply Fuses
Number:
2
Type:
T
Amps:
5 amps
Voltage:
250 volts
NOTE: Fuses on the power supply board are not externally accesible.
Replacement of fuses must be done by a qualified service technician
and the unit must be tested afterwards. Refer to the Service Manual
for fuse replacement.
Number:
2
Type:
H
Amps:
6.3 amps
Voltage:
250 volts