One of two icons appears in the upper left corner of the heart rate
numerical display box. These icons represent the gating signal
output—either waveform or pulse. The icon is the same color as the
SpO
2
waveform color when SpO
2
is selected for gating.
The gating function is user-selectable: OFF, ECG Waveform, ECG
Pulse, SpO
2
Waveform, or SpO
2
Pulse. The default is OFF.
1. Connect the cable appropriate for the scanner to the GATING
connection on the Veris patient monitor.
NOTE: The gating cable must be IEC 950 or IEC 60950
compliant.
2. Connect the MR scanner end of the cable to the scanner
interface receptacle.
NOTES:
To remove the gating cable, grip the cable connectors from both
ends and pull back firmly. Do not pull on the gating cable to
remove it from the monitor or scanner.
Inspect the gating cable and connections regularly for signs of
mechanical damage (e.g., cracked insulation, crimped, or bent
wires).
3. With the menu knob select the PARAMS menu.
4. At Gating select either SpO
2
Wave or SpO
2
Pulse.
5. Configure the scanner in accordance with the manufacturer’s
recommendations.
Waveform Output
Pulse Output
Содержание Veris 8600
Страница 1: ...Page i VerisTM 8600 Vital Signs Monitor Operation Manual 3010796 Revision 0 Date 10 04 PRELIMINARY...
Страница 10: ...This page intentionally left blank...
Страница 24: ......
Страница 34: ......
Страница 122: ......
Страница 142: ......
Страница 152: ......
Страница 164: ......
Страница 170: ......
Страница 190: ......
Страница 202: ......
Страница 212: ......
Страница 216: ......