MEDRAD
Veris
8600
2 —16
ALARMS Softkey
Figure 2-7: ALARMS Softkey Selected
This softkey allows access to all the parameter alarm settings. When
the menu is activated by pressing the menu knob, an alarm limit
settings window appears. The alarm window appears as the Adult,
Pediatric, or Neonate window as set in the third item Patient size.
Figure 2-8: Alarm Settings Window (Adult)
The pediatric and adult settings initially are identical, as factory
defaults, but can be adjusted independently and saved as desired.
Figure 2-9: Alarm Settings Window (Neonate)
ALARMS
PARAMS
DISPLAY
ZERO IP1
ADM/DIS
CONFIG
ZERO IP2
EXIT
Alarm Volume
5
ECG Lead Fail
MEDIUM
Patient size
Adult
HIGH
LOW
Heart Rate
150
40
SpO2
OFF
90
NIBP Systolic
200
50
NIBP Diastolic
100
30
NIBP Mean
150
50
Temperature 1 °
F
100.0 93.0
Temperature 2 °
F
100.0 93.0
Respiration
36
OFF
CO2 Ins mmHg
10
5
CO2 Exp mmHg
55
20
O2 Ins %
100
18
O2 Exp %
100
OFF
Apnea
20
seconds
Other Alarms
No Action
EXIT
Alarm Volume
5
ECG Lead Fail
MEDIUM
Patient size
Neonate
HIGH
LOW
Heart Rate
180
90
SpO2
OFF
90
NIBP Systolic
140
35
NIBP Diastolic
80
30
NIBP Mean
100
35
Temperature 1 °
F 100.0
93.0
Temperature 2 °
F 100.0
93.0
Respiration
60
14
CO2 Ins
mmHg
10
5
CO2 Exp
mmHg
55
20
O2 Ins
%
100
18
O2 Exp
%
100
OFF
Apnea
20 seconds
Other Alarm Setups
No Action
Содержание Veris 8600
Страница 1: ...Page i VerisTM 8600 Vital Signs Monitor Operation Manual 3010796 Revision 0 Date 10 04 PRELIMINARY...
Страница 10: ...This page intentionally left blank...
Страница 24: ......
Страница 34: ......
Страница 122: ......
Страница 142: ......
Страница 152: ......
Страница 164: ......
Страница 170: ......
Страница 190: ......
Страница 202: ......
Страница 212: ......
Страница 216: ......