MEDRAD
Veris
8600
A-4
Accidental Wetting
• Shock Hazard! The monitor is an AC powered device and an
immersed monitor presents a danger to anyone who handles
the device.
The action to be taken following accidental wetting of the equipment
is as follows:
1. Turn the power off! Disconnect the AC power cord from the
monitor, remote display, or power supply.
2. If monitoring a patient, transfer the patient to another monitor as
quickly as possible.
3. Use a clean, dry towel or cloth to remove the liquid from the
housing.
4. The monitor, remote display, or power supply should be
inspected by a service technician as soon as possible.
5. If the internal mechanism is saturated, allow the liquid to drain
out for 24 hours before shipping.
6. If liquid has entered the monitor, remote display, or power
supply, it needs to be dried and cleaned internally. Full testing is
required before the monitor, remote display, or power supply
can be used. Contact the MEDRAD Service Department as
soon as possible.
Time is critical! The longer any liquid remains in the monitor, remote
display, or power supply, the more damage it can do. It is important to
service the monitor, remote display, or power supply immediately
after any liquid is spilled into it.
WARNING
!
!
Содержание Veris 8600
Страница 1: ...Page i VerisTM 8600 Vital Signs Monitor Operation Manual 3010796 Revision 0 Date 10 04 PRELIMINARY...
Страница 10: ...This page intentionally left blank...
Страница 24: ......
Страница 34: ......
Страница 122: ......
Страница 142: ......
Страница 152: ......
Страница 164: ......
Страница 170: ......
Страница 190: ......
Страница 202: ......
Страница 212: ......
Страница 216: ......