2 —37
2 —Monitor Setup
PRINT Softkey
Figure 2-26: PRINT Softkey Selected
NOTE: The printer is only found in the remote display. Printer settings
may be adjusted from the main monitor or the remote display.
The PRINT softkey allows access to the printer settings window.
Figure 2-27: Print Configuration Window
If the internal thermal printer does not respond, check the CONFIG
menu. The Printer Device must be set to Internal Printer. If the
monitor returns to factory defaults the CONFIG setting returns to
Serial.
The setting Alarm Print causes the monitor to print all parameters
upon the activation of a new high or medium level alarm.
See “Printing and Data Ports” on page 12-1 for more information on
printing.
TABULAR PRINTING
Numerical values for all current parameters are printed.
GRAPHICAL PRINTING
If both Waveform 1 and 2 have been set to a physical parameter the
print out is a split dual waveform. If only one of the waveforms is
turned on, a single waveform is printed using the entire waveform
area. If both waveforms are turned off, the waveform area of the print
out is blank.
ALARMS
PARAMS
DISPLAY
ZERO IP1
ADM/DIS
CONFIG
ZERO IP2
EXIT
Print Type
Graphical
Alarm Print
OFF
BP Print
OFF
Interval Print
OFF
INTERVAL PRINT TYPE
TABULAR
Snapshot Size
6 Seconds
History Size
6 Seconds
Waveform 1
ECG II
Gain
x1.0
Waveform 2
Pleth
Gain
x1.0
Printer Speed
25 mm/sec
Содержание Veris 8600
Страница 1: ...Page i VerisTM 8600 Vital Signs Monitor Operation Manual 3010796 Revision 0 Date 10 04 PRELIMINARY...
Страница 10: ...This page intentionally left blank...
Страница 24: ......
Страница 34: ......
Страница 122: ......
Страница 142: ......
Страница 152: ......
Страница 164: ......
Страница 170: ......
Страница 190: ......
Страница 202: ......
Страница 212: ......
Страница 216: ......