2 —35
2 —Monitor Setup
Password Protection
Several settings on the CONFIG and the Enter Network Config
screens are password protected. These are functions that generally
should not be changed during use and their settings should be
adjusted by supervisory personnel, your MEDRAD representative, or
other qualified service personnel. These include; Standby Timeout,
Alarm Tone Warning, Line Frequency, Enter Service Mode, and Enter
Simulation Mode. Contact MEDRAD Service or your local
representative about service passwords.
Date Format
The date format can be set to display or print Day-Month-Year or
Month-Day-Year.
Time/Date Setting
The Time and Date are set in the CONFIG softkey window. Changing
the time and date while monitoring a patient does not affect the
accurate display of patient data, but it clears any recorded trend data.
Freeze Timeout
The FREEZE function can be set to freeze the waveform frame from
30 seconds to five minutes. Choosing OFF causes the FREEZE key
to hold the screen permanently until the FREEZE key is pressed
again. The FREEZE key also captures waveforms that are obscured
by dialog boxes and pop-up windows. When the FREEZE key is
pressed the waveforms are redrawn on the screen. The FREEZE
function forces an exit from the current pop-up window or dialog box.
Standby Timeout
The Standby Timeout can be set to disable the monitor from 5
minutes to 2 hours. Choosing OFF causes the STAND BY key to
disable the monitor permanently until the STAND BY key is pressed
again. This setting is password protected. The factory default value is
30 minutes.
Standby Tone
When set to ON the monitor produces a single low pitched beep
every minute during Standby Mode. When set to OFF there is no tone
to remind the user that the monitor is in Standby Mode.
Alarm Tone Warning
When set to ON the monitor produces a low pitched double beep
every two minutes during the permanent silence condition. When set
to OFF there is no tone to remind the user that the silence condition
is active.
NOTE: Set this safeguard according to your facility protocols and
according to local safety regulations for medical devices.
Print Device
The print device can be set to Internal Printer, Serial, or OFF.
• Set to Internal Printer to print data to the thermal printer on the
optional remote display.
• Set to Serial to print to an external printer or download to a
computer.
• Set to OFF to disable the printing feature.
See “Printing and Data Ports” on page 12-1 for more information.
Содержание Veris 8600
Страница 1: ...Page i VerisTM 8600 Vital Signs Monitor Operation Manual 3010796 Revision 0 Date 10 04 PRELIMINARY...
Страница 10: ...This page intentionally left blank...
Страница 24: ......
Страница 34: ......
Страница 122: ......
Страница 142: ......
Страница 152: ......
Страница 164: ......
Страница 170: ......
Страница 190: ......
Страница 202: ......
Страница 212: ......
Страница 216: ......