58 Troubleshooting
YM6000 Service Manual
Alarm
Messages
Check
Items
obtained.
It
is
possible
that
the
temperature
in
the
vicinity
of
the
sensor
is
extremely
low
(less
than
15.0°C)
or
extremely
high
(more
than
45.0°C).
Adjust
the
ambient
temperature
and
measure
again.
CAPNO:
Zero
calibration
range
error
Try
again
the
gas
calibration.
CAPNO:
Zero
calibration
signal
unstable
error
Try
again
the
gas
calibration.
CAPNO:
High
calibration
range
error
Try
again
the
gas
calibration.
CAPNO:
High
calibration
signal
unstable
error
Try
again
the
gas
calibration.
SYSTEM:
Low
battery.
Plug
the
AC
power
cord
to
the
AC
main
to
recharge
the
battery.
NIBP:
Retry,
Check
cuff/Patient
(C12)
Blood
pressure
could
not
be
measured
even
after
cuff
pressure
decreased.
It
is
possibly
because
pulse
was
not
strong
enough
for
measurement,
or
because
change
of
pulse
amplitude
could
not
be
obtained.
Check
whether
cuffs
are
not
wrapped
around
thick
clothing.
After
wrapping
cuffs
around
properly,
measure
again.
When
the
error
occurs
in
the
initial
measurement
in
continuous
mode,
the
second
measurement
will
start
unless
Stop
button
is
pressed.
NIBP:
Retry,
Cuff
excessive
artifact
(C13)
Measurement
failed
because
of
patient
movement
during
measurement.
Tell
the
patient
to
stay
still,
then,
measure
again.
When
it
occurs
in
the
initial
measurement
in
continuous
mode,
the
second
measurement
will
start
unless
Stop
button
is
pressed.
NIBP:
Retry,
Cuff
insufficient
pressure
(C14)
Measurement
failed
because
of
insufficient
pressurizing.
There
is
a
possibility
that
standard
cuff
pressure
might
be
detected
wrongly
due
to
noise,
patient
movement
or
external
vibrations.
Check
whether
cuffs
are
not
wrapped
around
thick
clothing,
whether
the
patient
is
moving
and
whether
there
is
external
vibration.
When
it
occurs
in
the
initial
measurement
in
continuous
mode,
the
second
measurement
will
start
unless
Stop
button
is
pressed.
NIBP:
Retry,
Cuff
irregular
pulses
(C15)
Blood
pressure
could
not
be
measured
because
oscillation
graph
was
not
normal.
There
is
a
possibility
that
patient
movement
or
external
vibration
interrupted
the
measurement.
Check
whether
a
patient
stays
still
and
cuffs
are
free
from
outside
vibrations,
then,
measure
again.
When
it
occurs
in
the
initial
measurement
in
continuous
mode,
the
second
measurement
will
start
unless
Stop
button
is
pressed.
NIBP:
Retry,
Cuff
motion
artifact
(C16)
Blood
pressure
could
not
be
measured
because
noises
interrupted
pulse
waveform
signal.
There
is
a
possibility
that
patient
movement
or
external
vibration
interrupted
the
measurement.
Confirm
the
patient
is
not
moving
and
the
cuff
is
free
of
external
vibration,
then,
measure
again.
When
it
occurs
in
the
initial
measurement
in
continuous
mode,
the
second
measurement
will
start
unless
Stop
button
is
pressed.
NIBP:
Retry,
Cuff
time
‐
out,
over
160
pulses
(C18)
Pulse
waveform
signal
more
than
160
beats
are
detected
during
measurement.
There
is
a
possibility
that
noise,
patient
movement
or
external
vibrations.
Confirm
the
patient
is
not
moving
and
the
cuff
is
free
of
external
vibration,
then,
measure
again.
NIBP:
Retry,
Cuff
pressure
failure
(C19)
Cuff
pressure
exceeded
more
than
300mmHg
during
measurement.
There
is
a
possibility
that
the
patient
moved
during
measurement
or
strong
pressure
from
outside
might
be
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...