![Mediana YM6000 Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/mediana/ym6000/ym6000_service-manual_1760618031.webp)
Performance Verification 25
YM6000 Service Manual
Air
Leakage
Test
The
air
leakage
test
verifies
the
integrity
of
the
pneumatic
system.
Figure
9.
Air
Leakage
Test
1.
Place
the
adult
cuff
with
a
rigid
PVC
vessel
(5
cm
diameter).
2.
Connect
the
cuff
to
the
NIBP
connector
on
the
monitor’s
front
panel
via
NIBP
cuff
hose.
3.
Rotate
the
trim
knob
to
select
NIBP
Test
in
the
service
menu,
and
then
press
the
trim
knob.
4.
Rotate
the
trim
knob
to
select
Air
Leakage
Test
,
and
then
press
the
trim
knob.
5.
The
result
will
be
displayed
on
the
monitor’s
screen
in
four
minutes.
6.
After
finishing
the
test,
press
the
trim
knob
to
select
Cancel
.
The
menu
box
will
disappear.
If
Cancel
is
selected
during
the
test
progressing,
the
test
will
stop
and
the
menu
box
will
disappear.
Pass/Fail
Results
It
passes
if
the
leakage
value
is
less
than
12mmHg/3minutes.
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...