Performance Verification 15
YM6000 Service Manual
Performance
Tests
The
battery
charge
and
battery
discharge
test
should
be
performed
before
monitor
repairs
whenever
the
battery
is
suspected
as
being
a
source
of
the
problems.
All
other
tests
may
be
used
following
repairs
or
during
routine
maintenance
(if
required
by
your
local
institution).
Before
performing
the
battery
discharge
test,
ensure
that
the
battery
is
fully
charged.
This
section
is
written
using
factory
defaults
set
as
power
‐
up.
If
your
institution
has
preconfigured
custom
defaults,
those
values
will
be
displayed.
Power
1.
Connect
the
monitor
to
AC
power
source
using
proper
power
cord.
2.
Verify
that
AC
indicator
is
lit.
3.
Press
Power
button
over
1
second
to
turn
on
the
monitor.
4.
Verify
that
the
monitor
is
turned
on
and
that
Power
on
indicator
is
lit
5.
After
the
monitor
operates
in
normal
mode,
disconnect
the
power
cord.
6.
Verify
that
Battery
status
icon
appears
on
the
screen
instead
of
lighting
AC
indicator.
7.
Press
Power
button
over
1
second,
and
then
verify
that
the
monitor
is
turned
off.
Battery
Charge
1.
Connect
the
monitor
to
AC
power
source
using
a
proper
power
cord.
2.
Verify
AC
indicator
is
lit.
3.
Charge
the
battery
fully
until
Battery
charging
status
indicator
A
and
B
is
no
more
flashing.
It
takes
for
at
least
12
hours.
4.
To
check
for
a
full
charge,
perform
the
procedure
in
paragraph
“
Battery
Discharge
”.
Note:
The
battery
may
require
a
complete
charge/discharge
cycle
to
restore
its
normal
capacity,
depending
on
its
previous
usage.
Battery
Discharge
1.
Disconnect
the
power
cord
from
the
monitor
with
fully
charge
battery.
2.
Turn
on
the
monitor
by
pressing
Power
button
over
1
second.
3.
Verify
Battery
status
icon
appears
at
bottom
of
the
screen
after
power
‐
on
self
‐
test
is
completed.
The
bar
in
battery
status
icon
should
be
filled,
indicating
battery
is
charged.
4.
Connect
SpO
2
simulator
to
monitor
via
the
SpO
2
extension
cable.
5.
Connect
NIBP
simulator
to
the
monitor
via
Mediana
hose.
6.
Set
SpO
2
simulator
as
follows:
SpO
2
of
90%
and
pulse
rate
of
60bpm.
7.
Set
NIBP
simulator
to
simulate
pressure
setting
of
120/80
mmHg
and
heart
rate
of
80bpm.
8.
Set
NIBP
Automatic
mode
interval
of
the
monitor
to
15
minutes.
9.
The
monitor
must
operate
for
1
hour
with
one
fully
charged
battery.
The
monitor
must
operate
for
at
least
15
minutes
before
the
monitor
powers
down
due
to
low
battery
condition.
10.
Verify
that
low
priority
audible
alarm
occurs
and
a
warning
message
“
low
battery
”
is
displayed
about
15
minutes
before
battery
fully
discharges.
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...