54 Troubleshooting
YM6000 Service Manual
Firmware
Download
If
the
error
code
appears
during
the
firmware
downloading,
take
the
following
action.
Table
16.
Firmware
Downloading
Error
Codes
Code
Description
Action
I.
1.
3
/
E.
1.
3
When
an
error
is
occurred
during
transferred
data.
Check
the
LAN
cable
and
settings
then
Retry
boot
download
I.
2.
3
/
E.
2.
3
When
an
error
is
occurred
during
transferred
of
data.
Check
the
LAN
cable
and
settings
then
Retry
boot
download
E.
1.
5
When
an
error
is
occurred
during
recorded
data.
Retry
boot
download
E.
2.
5
When
an
error
is
occurred
during
recorded
data.
Retry
main
download
E.
1.
7
When
the
data
is
broke
down
during
boot
download.
Retry
boot
download
E.
2.
7
When
the
data
is
broke
down
during
main
download.
Retry
main
download
Note:
If
the
Error
Code
still
appears,
take
monitor
out
of
service
and
contact
Mediana
Technical
Service
Department
for
advice
on
remedial
action
Technical
Alarm
Conditions
When
the
monitor
detects
the
technical
alarm
condition,
the
monitor
will
display
the
alarm
message
on
the
screen.
If
the
alarm
message
occurs,
follow
the
check
items
in
the
below
table
to
remove
the
alarm
condition
after
checking
the
alarm
message
in
the
alarm
message
area
or
informative
message
area.
Table
17.
Alarm
Messages
and
Check
Items
Alarm
Messages
Check
Items
NIBP:
Check
cuff
(C11)
Cuff
pressure
did
not
increase
enough
even
after
activating
the
pump
for
more
than
30
seconds
(adult).
There
is
a
possibility
that
a
cuff
hose
is
disconnected,
or
a
cuff
may
not
be
wrapped
around
an
arm.
Check
cuff
and
cuff
hose.
This
error
possibly
occurs
in
case
of
large
cuffs
that
are
wrapped
around
loosely.
When
the
error
still
occurs
even
after
checking
above,
there
is
a
possible
air
leakage
from
a
ruptured
cuff.
Replace
it
with
a
new
one.
NIBP:
Check
cuff
/
Patient
(C12)
Blood
pressure
could
not
be
measured
even
after
cuff
pressure
decreased.
It
is
possibly
because
pulse
was
not
strong
enough
for
measurement,
or
because
change
of
pulse
amplitude
could
not
be
obtained.
Check
whether
cuffs
are
not
wrapped
around
thick
clothing.
After
wrapping
cuffs
around
property,
measure
again.
When
the
error
occurs
in
the
initial
measurement
in
continuous
mode,
the
second
measurement
will
start
unless
Stop
button
is
pressed.
NIBP:
Cuff
excessive
artifact
(C13)
Measurement
failed
because
of
patient
movement
during
measurement.
Tell
the
patient
to
stay
still,
then,
measure
again.
When
it
occurs
in
the
initial
measurement
in
continuous
mode,
the
second
measurement
will
start
unless
Stop
button
is
pressed.
NIBP:
Cuff
insufficient
Measurement
failed
because
of
insufficient
pressurizing.
There
is
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...