30 Performance Verification
YM6000 Service Manual
To
perform
the
Two
Points
User
Calibration:
a.
Select
a
well
‐
ventilated
room
to
perform
the
calibration.
b.
Let
the
monitor
run
for
at
least
5
minutes
prior
to
calibration.
c.
For
the
“Sample
Zero
Reference
Gas”
step,
attach
the
CO
2
scrubber
to
the
water
trap
inlet
according
to
the
CO
2
scrubber
“Directions
for
Use”
(or
use
a
0%
medical
‐
grade
gas
source).
d.
Proceed
with
the
zero
gas
phase
of
the
calibration
as
defined
by
the
host
system.
Disconnect
CO
2
scrubber
(or
0%
gas
source)
from
the
water
trap
after
the
zero
gas
phase
and
prior
to
the
next
calibration
step.
e.
For
the
“Sample
Span
Reference
Gas”
phase
of
the
calibration,
provide
a
medical
‐
grade
gas
source
with
a
known
CO
2
concentration
between
8%
and
12%,
regulated
to
a
pressure
between
5
and
7
psi.
f.
Introduce
a
steady
stream
of
CO
2
gas,
Set
and
monitor
the
flow
rate
to
approximately
1
liter/minute
±10%).
To
verify
flow,
place
the
bleed
line
of
the
calibration
sample
line
in
a
glass
of
water.
If
bubbles
emerge,
gas
supply
is
sufficient.
If
no
bubbles
emerge,
there
is
an
insufficient
supply
of
gas.
g.
Connect
a
sample
line
from
the
sample
elbow,
or
Tee,
as
shown
above,
to
the
monitor’s
water
trap
inlet.
Make
sure
the
bleed
line
is
directed
away
from
the
monitor.
Allow
the
reference
gas
to
flow
for
at
least
one
minute.
h.
Sample
the
Span
Reference
Gas.
i.
After
the
Two
‐
Point
User
Calibration
steps
have
been
followed,
disconnect
the
monitor
from
the
gas
supply
and
test
setup.
j.
A
verification
may
be
performed
using
the
same
gas
delivery
set
up.
Verify
the
observed
CO
2
gas
reading
is
within
3
mmHg
or
10%,
whichever
is
greater,
of
the
CO
2
value
of
the
test
gas
supplied.
(If
not,
refer
the
user
to
the
appropriate
troubleshooting
information.)
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...