Performance Verification 21
YM6000 Service Manual
3.
Press
PROBE
button
on
PC.
4.
Verify
that
the
number
of
the
monitor
connections
to
PC
found
is
correct
5.
Press
EXIT
button
on
PC
to
close
“YM6000_Comm_Test.exe”.
Nurse
Call
Test
Perform
the
following
procedure
to
test
the
Nurse
Call.
The
nurse
call
connector
is
Dsub
‐
9,
located
on
the
monitor’s
right
panel,
identified
with
the
data
interface
symbol
(RS
‐
232).
1.
Connect
the
negative
lead
of
a
voltmeter
to
pin
5
and
positive
lead
to
pin
1
of
the
data
port
connector
on
the
right
panel
of
the
monitor.
Ensure
that
the
audible
alarm
is
not
silence
or
turned
off.
2.
Connect
the
SRC
‐
MAX
simulator
to
the
DOC
‐
10
sensor
cable.
3.
Connect
temperature
probe
to
temperature
input
port
on
YM6000.
4.
Turn
on
the
monitor
and
wait
for
the
monitor
to
complete
POST.
Note:
The
monitor
should
indicate
a
%SpO
2
alarm
of
75.
5.
Verify
an
output
voltage
at
pins
5
and
1
between
+5
to
+12
DC.
6.
Press
Alarm
silence/suspension
button.
With
no
active
audible
alarm,
the
output
voltage
at
pins
5
and
1
must
be
between
‐
5
to
‐
12
VDC.
This
verifies
the
RS
‐
232
Nurse
Call
function.
7.
With
the
instrument
in
an
alarm
condition,
use
a
digital
multmeter
(DMM)
to
verify
that
there
is
no
continuity
(1
mega
ohms
or
greater)
between
pins
8
and
9
and
that
there
is
continuity
(60
ohms
or
less)
between
pins
7
and
8.
8.
Press
the
SRC
‐
MAX
simulator
%SpO
2
button
to
change
the
%SpO
2
to
90.
9.
Use
a
DMM
to
verify
that
there
is
continuity
between
pins
8
and
9
and
that
there
is
no
continuity
between
pins
7
and
8.
This
verifies
the
solid
state
Nurse
Call
function.
Note:
The
pin
layouts
and
signal
descriptions
are
included
in
the
operator’s
manual.
For
the
detailed
information
regarding
the
Nurse
Call,
refer
to
Nurse
Call
Interface
section
of
the
operator’s
manual.
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...