34 Performance Verification
YM6000 Service Manual
Patient
Auxiliary
Current
This
test
measures
patient
auxiliary
current
in
accordance
with
IEC60601
‐
1,
clause
19,
for
Class
I,
type
CF
equipment.
The
applied
voltage
for
IEC60601
‐
1
the
voltage
is
264
volts,
50
to
60
Hz.
Patient
auxiliary
current
is
measured
between
each
ECG
test
lead
and
between
each
sensor
connection
for
all
possible
connections.
Note:
Keep
the
patient
test
cable
length
as
short
as
possible
during
the
leakage
test.
Note:
This
test
requires
the
same
test
cables
for
each
patient
connector
as
described
in
paragraph
“
Patient
Leakage
Current
”.
1.
Configure
the
electrical
safety
analyzer
as
recommended
by
the
electrical
analyzer’s
operating
instructions.
2.
Connect
the
monitor’s
AC
mains
power
cord
to
the
electrical
analyzer
as
recommended
by
the
electrical
analyzer’s
operating
instructions.
3.
Connect
the
patient
test
lead
combination
in
Table
7
to
the
appropriate
input
connector
on
the
electrical
analyzer.
4.
Turn
on
the
monitor.
5.
Perform
patient
auxiliary
current
test
per
Table
8
as
recommended
by
electrical
analyzer’s
operating
instructions.
6.
Repeat
the
patient
auxiliary
current
test
for
each
test
lead
combination
as
listed
in
Table
7
and
measure
each
patient
auxiliary
current.
Table
7.
Test
Lead
Combinations
First
Test
Lead
Second
Test
Lead
ECG
#1
(LA)
ECG
#2
(LL)
ECG
#1
(LA)
ECG
#
3
(RA)
ECG
#2
(LL)
ECG
#3
(RA)
ECG
#1
(LA)
Temperature
1/2
ECG
#2
(LL)
Temperature
1/2
ECG
#3
(RA)
Temperature
1/2
ECG
#1
(LA)
SpO
2
ECG
#2
(LL)
SpO
2
ECG
#3
(RA)
SpO
2
Temperature
1/2
SpO
2
Table
8.
Allowable
Leakage
Current
Test
Condition
Allowable
Leakage
Current
(microamps)
Normal
Condition
(NC)
10
SFC
Open
Supply
(OS)
50
SFC
Open
Earth
(SFC
OE)
50
Normal
Condition
RM
(NCRM)
10
SFC
Open
Supply
RM
(SFC
OSRM)
50
SRC
Open
Earth
RM
(SFC
OERM)
50
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...