28 Performance Verification
YM6000 Service Manual
IBP
Operation
1.
Connect
the
monitor
to
an
AC
power
source.
2.
Turn
on
the
monitor
by
pressing
the
Power
button.
3.
Connect
the
IBP
test
cable
to
the
IBP
simulator,
then
turn
on
the
IBP
simulator.
4.
Press
6,
zero
button
on
the
IBP
simulator,
then
press
enter
button
on
the
IBP
simulator.
5.
Connect
the
other
end
of
IBP
test
cables
to
IBP
channel
1
and
2
connectors
on
the
monitor’s
front
panel
after
completed
the
POST.
6.
The
monitor
will
display
IBP
waveform’s
baseline
on
the
screen.
7.
Select
IBP
waveform
menu
or
IBP
menu,
then
set
Pressure
Zero
Setting
,
to
Yes
to
perform
the
zero
calibration.
8.
If
zero
calibration
is
successful,
the
monitor
will
display
the
value
“0”
on
the
IBP
numerical
area.
9.
Press
the
5,
wavef
button
on
the
IBP
simulator,
then
press
enter
button
on
the
IBP
simulator.
10.
The
monitor
will:
‐
display
IBP
waveform
on
the
screen.
‐
display
systolic,
mean,
diastolic
measurement
values
on
the
IBP
numerical
area.
Note:
Initial
values
of
IBP
simulator
are
systolic
120mmHg,
diastolic
80mmHg
at
channel
1,
systolic
120mmHg,
diastolic
0mmHg
at
channel
2.
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...