Performance Verification 23
YM6000 Service Manual
13.
Set
ECG
Cable
Select
to
Auto
via
ECG
waveform
menu.
14.
Verify
that
Lead
Select
in
ECG
waveform
menu
displays
I,
II,
III.
15.
Disconnect
3
ECG
leads
and
connect
5
ECG
leads.
16.
Verify
that
Lead
Select
in
ECG
waveform
menu
displays
I,
II,
III,
aVR,
aVL,
aVF,
V
(Chest
Lead).
17.
Repeat
step
9
to
12.
18.
Turn
off
the
monitor.
Note:
The
accuracy
of
the
monitor
ECG
measurements
is
±3
bpm.
In
the
procedure,
add
the
tolerance
of
the
simulator
to
the
acceptable
range
of
readings.
NIBP
Operation
WARNING:
A
blood
pressure
cuff,
connected
to
the
monitor,
should
never
be
applied
to
a
human
subject
while
the
monitor
is
in
the
pressure
test
mode,
as
injury
could
result.
These
tests
verify
the
functionality
of
the
monitor’s
pneumatic
system.
The
monitor
must
be
placed
in
Service
menu
.
For
a
detailed
explanation
to
access
the
service
menu,
refer
to
Service
Menu
and
Factory
Default
Settings
section
.
1.
Rotate
the
trim
knob
to
select
NIBP
Test
in
the
service
menu,
and
then
press
the
trim
knob.
Note:
Before
accessing
the
NIBP
Test
mode,
ensure
that
current
patient
mode
is
proper
for
the
pneumatic
system
to
be
test.
You
can
set
Patient
mode:
Adult,
Pediatric
or
Neonatal
via
the
setup
menu.
Note:
In
the
NIBP
Test
mode,
no
function
button
will
have
no
effect
except
the
trim
knob.
All
the
tests
will
start
to
be
performed
by
pressing
or
rotating
the
trim
knob.
If
you
would
like
to
stop
the
test
during
test
progressing,
press
the
trim
knob.
Note:
Inflation
Speed
Test
and
Deflation
Speed
Test
are
intended
for
factory
use
only.
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...