Troubleshooting 53
YM6000 Service Manual
IBP
The
following
are
symptoms
of
problems
and
recommended
actions
relating
to
IBP
functions.
If
the
action
requires
replacement
of
a
PCB
assembly,
refer
to
Disassembly
Guide.
1.
The
ABP
and
CVP
window
does
not
exist
on
the
LCD.
1
‐
2
Please
go
to
the
Main
screen
of
setup
icon
and
set
the
display
window
to
6
‐
ch
Wave.
2.
The
IBP
values
are
erroneous.
2
‐
3
Check
the
IBP
accessories
connection
between
patient
and
each
end.
2
‐
4
Ensure
the
“Pressure
Zero
setting”
is
done
or
perform
the
“Pressure
Zero
setting”
at
IBP
menu.
2
‐
5
If
the
problem
persists,
replace
the
IBP
accessories.
2
‐
6
If
the
problem
persists,
replace
the
IBP
board.
EtCO
2
The
following
are
symptoms
of
problems
and
recommended
actions
relating
to
EtCO
2
functions.
If
the
action
requires
replacement
of
a
PCB
assembly,
refer
to
Disassembly
Guide.
1.
The
EtCO
2
is
not
displayed
at
LCD
1
‐
1
Please
remind
that
it
takes
3
minutes
to
warm
up
the
EtCO
2
module
after
turning
on
the
monitor.
1
‐
2
Please
go
to
the
Main
screen
of
setup
icon
and
set
the
display
window
to
6
‐
ch
Wave.
1
‐
3
Check
the
accessory
is
properly
connected
to
the
monitor
and
measurement
circuit.
1
‐
4
If
the
problem
persists,
replace
the
EtCO
2
accessories.
1
‐
5
If
the
problem
persists,
replace
the
EtCO
2
module.
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...