![Mediana YM6000 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/mediana/ym6000/ym6000_service-manual_1760618026.webp)
20 Performance Verification
YM6000 Service Manual
7.
Test
#3:
Record
speed
a.
Set
Record
speed
to
25
mm/s.
b.
Press
button
when
all
the
parameter
signals
display
normally.
c.
Verify
the
parameter
values
and
waveforms
are
printed
out
with
25
mm/s.
d.
Set
Record
speed
to
50
mm/s.
e.
Press
button
when
all
the
parameter
signals
display
normally.
f.
Verify
that
the
parameter
values
and
waveforms
are
printed
out
with
50
mm/s.
8.
Test
#4:
Wave
Record
Select
a.
Set
Wave
Record
Select
to
ECG1+ECG2
via
Setup
menu
.
b.
Press
Button
when
all
the
parameter
signals
display
normally.
c.
Verify
the
two
ECG
waveforms
are
printed
out.
d.
Repeat
this
test
for
other
selections.
9.
Test
#5:
Record
on
Alarm
a.
Set
Record
on
Alarm
to
On
via
Setup
menu
.
b.
Set
Heart
rate
of
ECG
simulator
to
30
bpm.
c.
Verify
”HR
Lower
limit
violated”
alarm
is
activated
and
the
parameter
values
and
waveforms
are
printed
out.
10.
Test
#6:
Auto
List
Record
a.
Set
Auto
List
Record
to
On
via
Setup
menu
.
b.
Set
Save
Time
Interval
to
0.5
minutes
via
Tabular
trend
menu.
c.
Verify
that
the
monitor
automatically
prints
out
the
data
after
stored
8
data
in
the
trend
memory.
Note:
If
there
is
no
printer
paper
left
or
printer
paper
places
improperly,
the
monitor
will
display
an
informative
message.
Network
Test
Perform
the
following
procedure
to
test
the
Network.
The
Network
connector
is
located
on
the
monitor’s
right
panel,
identified
with
the
Network
symbol.
CAUTION:
Do
not
change
any
other
settings
of
the
test
programs
while
performing
the
Network
test.
Note:
Network
Test
is
only
available
when
YM6000
monitor
is
equipped
with
TCP/IP
module
for
Central
Monitoring
System
(CMS).
Otherwise
this
test
does
not
need
to
perform.
Note:
Contact
Mediana
Technical
Service
Department
for
the
software
required.
1.
Connect
the
network
line
to
the
monitor,
then
turn
on
the
monitor.
2.
Run
“YM6000_Comm_Test.exe”
on
PC
connected
the
network
line
using
the
same
gateway
as
the
monitor.
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...